Z-0465-2016 Class II Terminated

Recalled by Boston Scientific Corp — San Jose, CA

Recall Details

Product Type
Devices
Report Date
December 30, 2015
Initiation Date
November 20, 2015
Termination Date
November 27, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1556 total - all models

Product Description

IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-10 ; Material Number: M004PM4790N40; Catalog number: PM4790N4; Cardiac: The IntellaTip MiFi XP Catheter is indicated for interruption of accessory atrioventricular conduction pathways associated with tachycardia, for treatment of AV nodal reentrant tachycardia, for treatment of atrial flutter tachycardia, and for creation of complete AV block in patients with a rapid ventricular response to an atrial arrhythmia  typically chronic, drug refractory atrial fibrillation.

Reason for Recall

During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSC's product risk documentation.

Distribution Pattern

Worldwide Distribution - US Nationwide in the state of: PA, FL, IN, MI, TN, AL, IL, KY, MI, TX, OH, MO, MA, MN, ID, AL, TN, VA, CA and internationally to Austria, Australia, Belgium, Switzerland, Germany, Great Britain, Greece, Hong Kong, Israel, Iran, Italy, Japan, Kuwait, Latvia, Malaysia, Netherlands, Saudi Arabia, Singapore, Thailand, and South Africa.

Code Information

Material Number: M004PM4790N40; Catalog number: PM4790N4; Lot numbers: 16743269, 16818653, 16818654, 16872007, 16993525, 17511087, 18159904, 18385792. Expiration Date Range: 9/1/2016 to 9/20/2018. .