Z-0480-2023 Class II Ongoing

Recalled by Ethicon, Inc. — Somerville, NJ

FDA device recall Z-0480-2023 was initiated by Ethicon, Inc. on Nov 16, 2022 and is designated Class II. Reason for recall: knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage The recall status is ongoing. Affected quantity: 7,560 eaches.

Recall details

Product Type
Devices
Report Date
Dec 21, 2022
Initiation Date
Nov 16, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7,560 eaches

Product description

PDS II (polydioxanone) Suture- VIOLET MONOFILAMENT 27IN 0 S/A CT-1 Product Code: Z340H For use in soft tissue approximation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery

Reason for recall

knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage

Distribution pattern

International (OUS) only: Canada, Japan, Australia and New Zealand

Code information

UDI-DI: 10705031060623 (individual unit); 30705031060634 (sales unit box) Lot: SCMSHK Exp. Date: February 28, 2027

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0480-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ethicon, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0480-2023. Report a data issue: hello@regdatalab.com.