Z-0490-2024 Class II Ongoing
FDA device recall Z-0490-2024 was initiated by Stanbio Laboratory, LP on October 2, 2023 and is designated Class II. Reason for recall: Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration… The recall status is ongoing. Affected quantity: 2,351.
Recall Details
- Product Type
- Devices
- Report Date
- December 13, 2023
- Initiation Date
- October 2, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,351
Product Description
STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.
Reason for Recall
Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of AZ, RI, MO, UT, AK, LA, TX, OH, GA, MN, NY, CA, NJ, IL, WA, KS, MI, NC, PA, MA, CO, KY, NM, FL, MD, VA, IN, AL, SD, MS, WV, WI, ID, MT, WY and the countries of Canada and Singapore.
Code Information
UDI-DI: 00657498000090, Lot/Expiration: 203044/2024-01-31, 209644/2024-06-30