Z-0518-2024 Class II Ongoing

Recalled by Riverpoint Medical, LLC — Portland, OR

FDA device recall Z-0518-2024 was initiated by Riverpoint Medical, LLC on Oct 25, 2023 and is designated Class II. Reason for recall: Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location… The recall status is ongoing. Affected quantity: 999 devices.

Recall details

Product Type
Devices
Report Date
Dec 20, 2023
Initiation Date
Oct 25, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
999 devices

Product description

Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix Knotless anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder.

Reason for recall

Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.

Distribution pattern

U.S. Nationwide distribution in the state of CA.

Code information

Product Code: 3911-7 14-520 UDI-DI Code: 00840277402172 Lot Number: 23080433 Product Code: 3911-7 14-521 UDI-DI Code: 00840277402189 Lot Number: 23081617

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0518-2024
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Riverpoint Medical, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0518-2024. Report a data issue: hello@regdatalab.com.