Z-0534-2013 Class II Terminated

Recalled by Arobella Medical, LLC — Minnetonka, MN

Recall Details

Product Type
Devices
Report Date
December 26, 2012
Initiation Date
November 5, 2012
Termination Date
March 4, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
99

Product Description

Arobella Medical 6mm Qurette used with the Arobella Medical AR1000 Wound Therapy System. Major components of the AR1000 Wound Therapy System include an ultrasonic generator, converter hand piece, done-shaped applicator tip (qurette), and a saline supply reservoir. Product Usage: The Arobella Medical AR1000 Wound Therapy System is interned for the selective dissection and fragmentation of tissue, wound debridement ( of acute and chronic wounds, burns, and diseased or necrotic tissue), and cleansing irrigation of the wound site for the removal of debris, exudates, fragments, and other matter.

Reason for Recall

This recall has been initiated because the Arobella Medical 6mm Qurette Q07 Lot Numbers 0001 and 0002 can fracture during use and may release a fragment or fragments. The 6 mm Qurette is a component of the AR1000 Quostic Wound Therapy System.

Distribution Pattern

Worldwide distribution - US (nationwide) and the states of AL, CA, ID, IN, IA, MI, MN, MS, NJ, NM, OH, PA. SC, TX, UT, VA, WI. including the country of SAUDI ARABIA.

Code Information

0001 and 0002