Z-0535-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- December 16, 2020
- Initiation Date
- November 12, 2020
- Termination Date
- January 12, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8 generators
Product Description
VNS Therapy SENTIVA DUO # 1000-D - Product Usage: indicated for the treatment of chronic or recurrent depression in patients who are experiencing a treatment-resistant or treatment- intolerant depressive episode.
Reason for Recall
During internal testing, it was found that upon a device reset, the Generator exhibits the incorrect Model Number when interrogated.
Distribution Pattern
International distribution in the countries of Austria and United Kingdom.
Code Information
Model Number - Model 1000-D Serial Numbers: 311298 310443 310068 310353 310531 310351 310380 310348