Z-0558-2018 Class II Ongoing

Recalled by Covidien LLC — North Haven, CT

Recall Details

Product Type
Devices
Report Date
February 14, 2018
Initiation Date
August 7, 2017
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 20cm, Material CA20L1

Reason for Recall

The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.

Distribution Pattern

Nationwide including PR, Canada, China

Code Information

UDI 10884521189546, Lot codes: S5LG003PX,S5MG016PX,S5MG018PX,S6AG004X,S6AG005X,S6AG007X,S6BG004X,S6BG011X,S6BG015PX,S6BG017X,S6CG002X,S6CG008X,S6CG014X,S6CG015X,S6CG016X,S6DG002X,S6DG005X,S6DG007X,S6DG010X,S6EG004X,S6EG006X,S6EG008X,S6FG005X,S6FG008X,S6GG007X,S6GG010X,S6HG001X,S6HG003PX,S6HG008PX,S6HG009PX,S6HG012X,S6JG001X,S6KG003X,S6KG005PRX,S6KG005PX,S6KG009X,S6KG016X,S6LG002X,S6MG003X,S6MG018X,S7AG003X,S7BG012PCX,S7CG004X,S7CG006X,S7CG008PCX,S7CG010PCX,S7CG017X,S7DG002X,S7DG005X,S7EG006X,S7FG010PX