Z-0574-2020 Class II Terminated

Recalled by Abbott Gmbh & Co. KG — Wiesbaden

Recall Details

Product Type
Devices
Report Date
December 4, 2019
Initiation Date
November 7, 2019
Termination Date
August 31, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
83 kits

Product Description

ARCHITECT HAVAB-G Calibrator, REF 6L27-01

Reason for Recall

Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.

Distribution Pattern

US in the states of: AZ, CA, CO, CT, FL, KY, LA, MA, MD, MI, MN, MO, NJ, NY, OH, OR, RI, SC, TN, TX, WA, WV

Code Information

lot number 08219BE00