Z-0574-2026 Class II Ongoing

Recalled by Sterilmed, Inc. — Plymouth, MN

Recall Details

Product Type
Devices
Report Date
November 26, 2025
Initiation Date
October 8, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4

Product Description

Brand Name: Sterilmed Reprocessed Imaging Catheter Product Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135936 Software Version: N/A Product Description: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR USE WITH SIEMENS SYSTEM Component: N/A

Reason for Recall

Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.

Distribution Pattern

United States Only: Kentucky, Maryland, Massachusetts, Mississippi, New Jersey, Pennsylvania, Texas, Utah

Code Information

Lot Code: GTIN: 10888551004732 LOT# SERIAL#: 2188147 131065 2207154 916179 2220232 1050886 2234795 1170283