Z-0580-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- December 17, 2014
- Initiation Date
- November 20, 2014
- Termination Date
- August 11, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 184 units
Product Description
IntellaMap Orion High Resolution Mapping Catheter; Material number: M004RC64S0; Catalog number RC64S; Is indicated for electrophysiological mapping (recording or stimulating only) of the cardiac structures of the heart.
Reason for Recall
Some IntellaMap Orion High Resolution Mapping Catheters are exhibiting a failure mode in which a glue bond on the Deployment Interlock Button has been observed to fail in the field.
Distribution Pattern
Worldwide Distribution -- US, including the states of MA, AL, NC, and OK; and, the countries of Germany, France, and Great Britain.
Code Information
Lot numbers: 17075550, 17075981, 17083913, 17088826, 17094081, 17101668, 17208445, 17218467, 17251661, 17258617, 17274937, 17292792, and 17308586. Expiry Dates from June 25, 2015 to September 24, 2015.