Z-0592-2018 Class II Terminated

Recalled by Cepheid — Sunnyvale, CA

Recall Details

Product Type
Devices
Report Date
February 21, 2018
Initiation Date
October 9, 2017
Termination Date
February 13, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,586 kits (49,400 units)

Product Description

Xpert CT/NG Catalog GXCT/NG-120 Batch 1000049568, Catalog GXCT/NG-10 Batch 1000053075, Catalog GXCT/NG-10 Batch 1000056156 and Catalog GXCT/NG-10 Batch 1000056157. Is a qualitative in vitro real-time PCR test.

Reason for Recall

Cepheid has received reports that some of our customers have experienced a higher number of temperature drift errors than would typically be expected when using this test and its intended use specimen types. Investigation of the issue also revealed an occasional probability of false negative results which remains consistent with package insert claims.

Distribution Pattern

Worldwide Distribution -- USA, Puerto Rico, and Haiti.

Code Information

Lot numbers: 25718, 25719, 25725, and 25727.