Z-0601-2024 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics Inc — Norwood, MA

Recall Details

Product Type
Devices
Report Date
December 27, 2023
Initiation Date
October 19, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10,183 devices

Product Description

epoc NXS Host Blood Analysis System, Siemens Material Numbers 11413497 (EU), 11413475 (US), 11413506 (CA), 11413517 (MX), 11413518 (ROW), 11413498 (JP), 11413583 (CN, KR), 11413879 (IN) with software versions from 4.9.4 to the current version 4.11.11

Reason for Recall

Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing erroneous results to be printed. When the results are run on the epoc NXS Host and Delete Blood Tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. The worst possible outcome may result in a failure to diagnose hyperkalemia, due to an erroneous depressed potassium level or inappropriate treatment to hyperglycemia due to erroneous glucose reading.

Distribution Pattern

US Nationwide. Global Distribution.

Code Information

UDI-DI: 00630414605760 00630414606095 00630414605821 00630414605838 00630414605678 00630414605814 00630414612447 00630414631028; All serial numbers.