Z-0630-2016 Class I Terminated

Recalled by Arkray Factory USA, Inc. — Minneapolis, MN

FDA device recall Z-0630-2016 was initiated by Arkray Factory USA, Inc. on Dec 18, 2015 and is designated Class I. Reason for recall: One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected… The recall status is terminated (terminated Jan 4, 2017). Affected quantity: 88 boxes (25 foil packaged test strips per box).

Recall details

Product Type
Devices
Report Date
Jan 27, 2016
Initiation Date
Dec 18, 2015
Termination Date
Jan 4, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
88 boxes (25 foil packaged test strips per box)

Product description

arkray SPOTCHEM II Glucose Reagent test strips, Item No. 77101, MEDTEST DX Catalog # 77424. For the in vitro quantitative determination of glucose in human plasma from whole blood samples.

Reason for recall

One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.

Distribution pattern

Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.

Code information

Lot Number EA4M78

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0630-2016
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0630-2016. Report a data issue: hello@regdatalab.com.