Z-0640-2019 Class II Terminated

Recalled by Advanced Instruments, LLC — Norwood, MA

Recall Details

Product Type
Devices
Report Date
December 26, 2018
Initiation Date
February 7, 2018
Termination Date
September 24, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity

Product Description

GloCyte RBC Reagent - Product Usage: The GloCyte Automated Cell Counter System is intended for use by trained healthcare professionals in clinical laboratories to provide quantitative determination of fluorescence labeled total nucleated cells and erythrocytes in cerebrospinal fluid collected from adult and pediatric patients. The GloCyte Low and High Level Controls are assayed hematology controls designed to monitor the performance of the GloCyte Automated Cell Counter System. Assayed parameters include total nucleated cells and erythrocytes.

Reason for Recall

RBC Reagent failure. High fluorescent counts during RBC Reagent + Diluent QC check that will result in failed QC check, preventing user from testing patient samples for RBC.

Distribution Pattern

US Nationwide distribution in the states of IN, WA, NC, CA, GA, CO.

Code Information

Part number 111011, Lot # R1513