Z-0646-2017 Class II Terminated

Recalled by Lsi Medience Corporation — Tokyo, N/A

FDA device recall Z-0646-2017 was initiated by Lsi Medience Corporation on Aug 16, 2016 and is designated Class II. Reason for recall: Two complaints were received about decrease in sensitivity (control value and/or calibration value) of the product. The recall status is terminated (terminated Feb 27, 2017). Affected quantity: 42 boxes (each containing 60 test cartridges/one kit contains 10 reagent cartridge trays and one tray has 6 cartridges).

Recall details

Product Type
Devices
Report Date
Nov 30, 2016
Initiation Date
Aug 16, 2016
Termination Date
Feb 27, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 boxes (each containing 60 test cartridges/one kit contains 10 reagent cartridge trays and one tray has 6 cartridges).

Product description

Pathfast D-Dimer, Reference number: PF1051-KUS

Reason for recall

Two complaints were received about decrease in sensitivity (control value and/or calibration value) of the product.

Distribution pattern

US Distribution to the states of : AL, FL, ID and MA.

Code information

Lot T528

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0646-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Lsi Medience Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.