Z-0650-2013 Class II Terminated

Recalled by California Medical Laboratories Inc — Costa Mesa, CA

Recall Details

Product Type
Devices
Report Date
January 16, 2013
Initiation Date
November 2, 2012
Termination Date
March 4, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
S112468N=1000 units, S120196N=1400 units, & S120197N=1200 units

Product Description

Rigid Tip Suction Wand Non sterile bulk devices (S112468N, S120196N, and S120197N) were sold to Terumo UK to be included in a kit under Terumo as manufacturer's name. The Suction Wands are indicated for use to remove excess fluid from the surgical field.

Reason for Recall

California Medical Laboratories, Inc. has identified an issue with the Rigid Tip Suction Wand. Due to insufficient adhesive application, the distal tip has the potential to come loose from the wand and fall into the patient's chest cavity.

Distribution Pattern

Worldwide Distribution - USA including AZ, FL, CA, LA, CO, and MS. Internationally to Canada, Israel, Greece, Hong Kong, India, Ireland, Jordan, Lebanon, Malaysia, Netherlands, Russia, Saudi Arabia, South Africa and Spain.

Code Information

Catalog No: SU-12202NS Lot No: S112468N, S120196N, S120197N