Z-0664-2019 Class II Terminated

Recalled by Abbott Ireland Diagnostics Division — Co. Longford

Recall Details

Product Type
Devices
Report Date
January 2, 2019
Initiation Date
November 21, 2018
Termination Date
April 1, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
678 units

Product Description

Alinity i TSH Reagent Kit, List Number 07P4830

Reason for Recall

Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.

Distribution Pattern

AR, FL, GA, MD, OK, VA

Code Information

UDI/Lot Numbers: (01)00380740131159 (17)190710(10)90061UI00 / 90061UI00 (01)00380740131159 (17) 190827(10) 91526UI00 / 91526UI00 (01)00380740131159 (17)190924(10)92291UI00 / 92291UI00