Z-0678-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- December 18, 2019
- Initiation Date
- December 3, 2019
- Termination Date
- September 18, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11 distributed in the US
Product Description
Maquet Cardiopulmonary (MCP)/Getinge ROTAFLOW Drive, Model Number 701022161. The device is a component of the ROTAFLOW System Console.
Reason for Recall
A loose ODU plug / Coaxial cable connection may result in fluctuating flow values on the Console.
Distribution Pattern
The products were distributed to the following US states: ME, NV, NY, OH, PA, TX, and VA.
Code Information
UDI: 04037691258997; Serial Numbers 9348936 910114412 910114420 910114429 910114430 910114432 910114435 910114438 910117243 91064722 91067680