Z-0702-2017 Class II Terminated

Recalled by Thayer Intellectual Property, Inc. — Walnut Creek, CA

Recall Details

Product Type
Devices
Report Date
December 14, 2016
Initiation Date
November 4, 2016
Termination Date
December 13, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
339 devices

Product Description

MANOS EX (CAT#TY-12-101) Product Usage: Carpal Tunnel Release: as a tool for tissue release of the transverse carpal ligament in patients with Carpal Tunnel Syndrome who fail conservative therapy. Plantar Fasciotomy: as a tool for tissue recession in patients with plantar fasciitis who fail conservative therapy.

Reason for Recall

The firm failed to notify current customers of IFU update/warning. During a recent FDA inspection, it was discovered that Thayer Intellectual Property, Inc. had revised their IFU to include a warning not to bend, deform or otherwise modify any part of the device. The warning was added as a result of a MAUDE report filed in December 2014 where a piece of the MANOS EX broke during use and was left in the patient's hand. The patient suffered a tendon injury and an additional surgery was required to remove the broken piece of the device and repair the tendon. FDA inspection found that while the firm had updated their IFU, they had not notified customers who had been shipped the device with the older version that did not include the warning. Risk: It is important for MANOS EX users to be aware that bending, deforming, or modifying the device could result in device breakage which could impact patient safety and device performance.

Distribution Pattern

US Nationwide Distribution in the states of CA, TX. MD, MI,VA, DC, UT, CT, ID, SD, IN, and GA.

Code Information

LOT 1055337, Expiry 4/01/2015; LOT 1071800. Expiry 5/1/2017; LOT 1102580; Expiry 8/1/2017; LOT 1154179; Expiry 12/01/2017.