Z-0704-2024 Class II Ongoing

Recalled by BioFire Diagnostics, LLC — Salt Lake City, UT

Recall Details

Product Type
Devices
Report Date
January 17, 2024
Initiation Date
November 27, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
57 Kits

Product Description

The BIOFIRE Respiratory Panel 2.1 (RP2.1) is a PCR-based multiplexed nucleic acid test intended for use with the BIOFIRE FILMARRAY 2.0 or BIOFIRE FILMARRAY Torch Systems.

Reason for Recall

A manufacturing event may contribute to elevated rates of false negative results while using gastrointestinal panel and respiratory panel, which could lead to inappropriate antibiotic treatment, no treatment, delay of treatment, and may provide false assurance to someone with a true infection, delayed identification of an outbreak, and may result in additional transmission.

Distribution Pattern

US: NC, MO, NE, CT, IL, TN, NJ, OH, MA, CA, FL, UT, IA, KS, TX, MD

Code Information

UDI-DI: 00815381020529. Kit Lot/Pouch Lot/Expiration: 1667923/2X2423/June 26, 2024; 1890223/2YHT23/July 26, 2024