Z-0721-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 1, 2025
- Initiation Date
- November 6, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 318 kits
Product Description
Medline brand, medical procedure convenience kits, labeled as: 1) MVFF/FIBU #54-RF , SKU DYNJ21927R; 2) DIEP FLAP, SKU DYNJ910120A
Reason for Recall
Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods
Distribution Pattern
US Nationwide. Canada.
Code Information
UDI-DI: DYNJ21927R, 10195327104733(each), 40195327104734(case); Lot Number: 23IBF547 23GBT501 23FBU360 23EBU929 23EBL624 23EBA114 23CBN730 23ABF393 22JBK203 22IBN858 22HBO301 22GBC794 22CBI774; UDI-DI: DYNJ910120A, 10195327689872 (each), 40195327689873 (case), Lot Numbers: 24HMK254