Z-0722-2024 Class II Ongoing

Recalled by Cook Medical Llc — Bloomington, IN

FDA device recall Z-0722-2024 was initiated by Cook Medical Llc on Dec 7, 2023 and is designated Class II. Reason for recall: The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an… The recall status is ongoing. Affected quantity: 14 US; 101 OUS.

Recall details

Product Type
Devices
Report Date
Jan 24, 2024
Initiation Date
Dec 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14 US; 101 OUS

Product description

MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

Reason for recall

The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.

Distribution pattern

Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM

Code information

UDI-DI 00827002444294 Lots AM109365 AM109363 AM109356 AM111934 AM111932 AM111933 AM112041 AM112042 AM112039 AM112040 AM112043 AM112071 AM112072 AM112316

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0722-2024
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Cook Medical Llc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0722-2024. Report a data issue: hello@regdatalab.com.