Z-0728-2018 Class III Terminated

Recalled by Perkinelmer Life Sciences, Inc. — Boston, MA

FDA device recall Z-0728-2018 was initiated by Perkinelmer Life Sciences, Inc. on Aug 2, 2017 and is designated Class III. Reason for recall: We have become aware through internal testing that JANUS and Janus G3 Automated Workstations were failing production final acceptance testing due to a Horizontal Motion Error message displayed on… The recall status is terminated (terminated Jul 31, 2018). Affected quantity: 7.

Recall details

Product Type
Devices
Report Date
Mar 7, 2018
Initiation Date
Aug 2, 2017
Termination Date
Jul 31, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7

Product description

JANUS & JANUS G3 Automated Workstation is an automated programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows.

Reason for recall

We have become aware through internal testing that JANUS and Janus G3 Automated Workstations were failing production final acceptance testing due to a Horizontal Motion Error message displayed on the system. Further investigation concluded that two printed circuit boards (PCB) used in MDT models of Janus and Janus G3 were manufactured with the incorrect resistor values. This anomaly affects component part number 7101640 (PCB-Quad Motor Controller MDT) and part number 7101647 (PCB-Quad Motor Controller Gripper). This anomaly could result in the Horizontal Motion Error being displayed and result in instrument down time. Until PerkinElmer s corrective measures are completed, we request you; ¿ Be aware that there is potential for the Horizontal Motion Error message to display which could result in instrument down time. ¿ Follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. ¿ If you have additional questions, contact PerkinElmer Technical Support for further assistance (800) 762-4000.

Distribution pattern

Nationwide

Code information

AJM4G01, AJL8001, AJI4G01, AJL8M01, CJM8M01, CJL8M01, CJL8001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0728-2018
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Perkinelmer Life Sciences, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0728-2018. Report a data issue: hello@regdatalab.com.