Z-0738-2016 Class II Terminated

Recalled by Horiba Instruments Inc — Irvine, CA

Recall Details

Product Type
Devices
Report Date
February 10, 2016
Initiation Date
January 11, 2016
Termination Date
June 7, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
216

Product Description

The ABX PENTRA Cholesterol CP ref. A11A01634 is an in vitro diagnostic assay for quantitative determination of cholesterol in human serum and plasma based on an enzymatic photometric test (Trinders reaction). The assay is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.

Reason for Recall

N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.

Distribution Pattern

U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.

Code Information

All Lots