Z-0738-2016 Class II Terminated

Recalled by Horiba Instruments Inc — Irvine, CA

FDA device recall Z-0738-2016 was initiated by Horiba Instruments Inc on Jan 11, 2016 and is designated Class II. Reason for recall: N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol… The recall status is terminated (terminated Jun 7, 2016). Affected quantity: 216.

Recall details

Product Type
Devices
Report Date
Feb 10, 2016
Initiation Date
Jan 11, 2016
Termination Date
Jun 7, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
216

Product description

The ABX PENTRA Cholesterol CP ref. A11A01634 is an in vitro diagnostic assay for quantitative determination of cholesterol in human serum and plasma based on an enzymatic photometric test (Trinders reaction). The assay is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.

Reason for recall

N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.

Distribution pattern

U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.

Code information

All Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0738-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Horiba Instruments Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Horiba Instruments Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Horiba Instruments Inc →

Source: openFDA record for Z-0738-2016. Report a data issue: hello@regdatalab.com.