Z-0741-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 28, 2022
- Initiation Date
- November 8, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 140
Product Description
InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA, MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA.
Reason for Recall
Diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error message may be displayed indicating the app has stopped or isn't responding. After the crash, a present dose reminder is not displayed, which may lead to delayed insulin therapy and potentially resulting in hyperglycemia.
Distribution Pattern
US Nationwide distribution including in the states of CA, NY, MT, OH, IN, MN, PA, CT, TX, WA, TN, SD, MI, FL, OR, NC, KY, WI, MO, WV, NV, SC, AZ, AR, MA, LA, VT, GA.
Code Information
InPen Software App for Android, versions 5.6.1.1 and 5.6.2 with InPen System REF/UDI-DI: MMT-105ELBLNA/10862088000313, MMT-105ELGYNA/10862088000320, MMT-105ELPKNA/10862088000337, MMT-105NNBLNA/10862088000344, MMT-105NNGYNA/10862088000351, MMT-105NNPKNA/10862088000368.