Z-0741-2023 Class II Ongoing

Recalled by Companion Medical, Inc. — San Diego, CA

Recall Details

Product Type
Devices
Report Date
December 28, 2022
Initiation Date
November 8, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
140

Product Description

InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA, MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA.

Reason for Recall

Diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error message may be displayed indicating the app has stopped or isn't responding. After the crash, a present dose reminder is not displayed, which may lead to delayed insulin therapy and potentially resulting in hyperglycemia.

Distribution Pattern

US Nationwide distribution including in the states of CA, NY, MT, OH, IN, MN, PA, CT, TX, WA, TN, SD, MI, FL, OR, NC, KY, WI, MO, WV, NV, SC, AZ, AR, MA, LA, VT, GA.

Code Information

InPen Software App for Android, versions 5.6.1.1 and 5.6.2 with InPen System REF/UDI-DI: MMT-105ELBLNA/10862088000313, MMT-105ELGYNA/10862088000320, MMT-105ELPKNA/10862088000337, MMT-105NNBLNA/10862088000344, MMT-105NNGYNA/10862088000351, MMT-105NNPKNA/10862088000368.