Z-0742-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 1, 2025
- Initiation Date
- October 28, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 18
Product Description
Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
Reason for Recall
Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.
Distribution Pattern
US: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC
Code Information
UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1506, CRB-1508, CRB-1516, CRB-1517, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1527, CRB-1528, CRB-1529, CRB-1530, CRB-1532,