Z-0778-2024 Class II Ongoing

Recalled by Accure Acne, Inc — Boulder, CO

Recall Details

Product Type
Devices
Report Date
January 31, 2024
Initiation Date
December 6, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
22

Product Description

Accure Laser System, Model: PFMS00004

Reason for Recall

Laser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.

Distribution Pattern

US distribution to states of: CO, CA, CT, NY, NJ, MN, LA, OH, MA, TX

Code Information

UDI-DI: 08059173392230, Serial Numbers: LAD0000-0004, LAD0000-0014, PROT01-0720, LAD0000-0012, LAD0000-0005, LAD2216-0623, LAD2356-0723, LAD0000-0003, PROTO07-1120, LAD2076-0623, LAD0000-0011, LAD0000-0008, LAD0000-0010, LAD2357-0723, LAD2075-0623, LAD2232-0723, LAD0000-0013, LAD0000-0009, LAD0000-0006, LAD0000-0001, PROT06-1120, LAD2507-0723