Z-0790-2020 Class II
Recall Details
- Product Type
- Devices
- Report Date
- January 15, 2020
- Initiation Date
- November 28, 2018
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 274 devices
Product Description
ImplantDirect InterActive Precision IO Scan Adapter, Part #6534-09PT, Platform: 3.4mm, Non-Sterile, Rx. The firm name on the label is Implant Direct Sybron Manufacturing LLC, Thousand Oaks, CA.
Reason for Recall
The peek part was loose on the titanium base of the device which could result in it falling into a patient's mouth during a procedure resulting in aspiration or ingestion.
Distribution Pattern
Distribution was made to AL, AZ, CA, FL, GA, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, and VA. There was no government/military distribution. Foreign distribution was made to Canada, Armenia, Australia, Denmark, France, Germany, Great Britain, Iceland, Italy, Lithuania, Netherlands, Poland, Spain, Sweden, and Switzerland.
Code Information
Lot numbers 104479 and 104963.