Z-0811-2019 Class I Ongoing

Recalled by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) — Mounds View, MN

FDA device recall Z-0811-2019 was initiated by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) on January 17, 2019 and is designated Class I. Reason for recall: A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode wit… The recall status is ongoing. Affected quantity: 2040 units.

Recall Details

Product Type
Devices
Report Date
February 20, 2019
Initiation Date
January 17, 2019
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2040 units

Product Description

Medtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01

Reason for Recall

A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality

Distribution Pattern

Wordlwide

Code Information

Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01, GTIN: 00643169708969, 00643169708983, 00643169709072, All Serial Numbers