Z-0814-2024 Class II Ongoing
FDA device recall Z-0814-2024 was initiated by St. Jude Medical, Atrial Fibrillation Division, Inc. on Dec 18, 2023 and is designated Class II. Reason for recall: When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model… The recall status is ongoing. Affected quantity: 483 units.
Recall details
- Product Type
- Devices
- Report Date
- Jan 31, 2024
- Initiation Date
- Dec 18, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 483 units
Product description
Reason for recall
Distribution pattern
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0814-2024
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for St. Jude Medical, Atrial Fibrillation Division, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.