Z-0825-2022 Class II Terminated

Recalled by Dignitana, Inc. — Dallas, TX

FDA device recall Z-0825-2022 was initiated by Dignitana, Inc. on Nov 5, 2021 and is designated Class II. Reason for recall: Instructions for Use (IFUs) updated to include use of a headband is now recommended for all patients and instructions for the application of the headband. Additionally, IFU updated to include warnings… The recall status is terminated (terminated Oct 29, 2024). Affected quantity: 499 dignicaps.

Recall details

Product Type
Devices
Report Date
Mar 30, 2022
Initiation Date
Nov 5, 2021
Termination Date
Oct 29, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
499 dignicaps

Product description

DigniCap Delta Scalp Cooling System, Product Number: 900-1001, Labeling: The DigniCap Scalp Cooling System User Manual (PN: 0P1DDCD01M-EN_C / IFU 20190325-06-EN), DigniCap Delta Scalp Cooling Wrap (LAB 20190221-05-EN), DigniCap Delta Small Cooling Wrap (LAB 20200616-01-EN, Quick Reference Guide DigniCap Delta (IFU-20190327-02-EN), The DigniCap Scalp Cooling System Training Guide- Global (TR-20200104-05-EN), The DigniCap Scalp Cooling System Training Guide TR-2019-0801-05-US, DigniCap Delta Procedural Expectations- Global (TR-20191028-03-EN), DigniCap Delta Procedural Expectations TR- 20190721-02-US, DigniCap Delta Small Wrap Protocol Global(TR-20200728-02-EN), DigniCap Delta Small Wrap Protocol TR-20200714-01-US, DigniCap Delta Training Tips Global (TR-20200420-04-EN), DigniCap Delta Training Tips TR-20191202-05-US,

Reason for recall

Instructions for Use (IFUs) updated to include use of a headband is now recommended for all patients and instructions for the application of the headband. Additionally, IFU updated to include warnings for risk of condensation build up depending on environmental conditions with instructions.

Distribution pattern

US Distribution to states of: LA, NC, NJ, FL, WI, CA, NY, KS, GA, SC, CT and KY.

Code information

All manufactured units lots/serial numbers: A65860-A73244

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0825-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Dignitana, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0825-2022. Report a data issue: hello@regdatalab.com.