Z-0831-2025 Class I
Recall Details
- Product Type
- Devices
- Report Date
- January 22, 2025
- Initiation Date
- December 11, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 845 units
Product Description
nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System
Reason for Recall
The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
Distribution Pattern
US Nationwide Distribution and the country of Hong Kong
Code Information
Lot Numbers: 22697, 22698, 22699, 22700, 22701, 22702, 22703, 22704, 22705, 22706, 22707, 22708, 22721, 22722, 22723, 22724, 22733, 22734, 22735, 22736, 22737, 22738, 22739, 22740.