Z-0850-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 6, 2022
- Initiation Date
- February 11, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 653 (US 149, OUS 504)
Product Description
RayStation 4-11B, and RayPlan 1, 2, 7-11B including some service packs: RayStation 4.0 (4.0.0.14) to RayStation 5 Service Pack 2 (5.0.2.35), RayStation 5 Service Pack 3 (5.0.3.17), RayStation 6/RayPlan 2 (6.0.0.24), RayStation 6/RayPlan 2 Service Pack 1 (6.1.1.2), RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7), RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6), RayStation/RayPlan 7 (7.0.0.19), RayStation/RayPlan 8A (8.0.0.61), RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) Software versions: 4.0.0.14, 4.0.3.4, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.7.6.7, 4.9.0.42, 5.0.1.11, 5.0.2.35, 5.0.3.17, 6.0.0.24, 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 10.1.1.54, 11.0.0.951, 11.0.1.29, 11.0.3.116 or 12.0.0.932
Reason for Recall
Incorrect source to skin or surface distance (SSD) calculation may result in dose error when using RayStation system.
Distribution Pattern
US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CT, DC, DE, FL, GA, HI, IA, ID, IL, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI.
Code Information
UDI # 07350002010020, 07350002010013, 07350002010082, 07350002010075, 07350002010242, 07350002010068, 07350002010112, 07350002010136, 07350002010129, 07350002010204, 07350002010235, 07350002010174, 07350002010266, 07350002010297, 07350002010303, 07350002010365, 07350002010310, 07350002010471, 07350002010389, 07350002010433, 07350002010440, 07350002010426