Z-0867-2022 Class II Ongoing
FDA device recall Z-0867-2022 was initiated by Shimadzu Medical Systems on Jan 10, 2022 and is designated Class II. Reason for recall: It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation. The recall status is ongoing. Affected quantity: 25 systems.
Recall details
- Product Type
- Devices
- Report Date
- Apr 13, 2022
- Initiation Date
- Jan 10, 2022
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 25 systems
Product description
Reason for recall
Distribution pattern
U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A