Z-0867-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 13, 2022
- Initiation Date
- January 10, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 25 systems
Product Description
MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
Reason for Recall
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
Distribution Pattern
U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A
Code Information
Model: FLUOROspeed X1 Generator Number: D150-40 Serial Numbers: MQ493E19C001 MQ493E3B8002 MQ493E3B9001 MQ493E1B4001 MQ493E1A1003 MQ493E1A3001 MQ493E1B5001 MQ493E3B8001 MQ493E1A5001 MQ493E1A1005 MQ493E1A2001 MQ493E1A4001 MQ493E1B6002 MQ493E1A8001 MQ493E1B4002 MQ493E1A4003 MQ493E1A4002 MQ493E1B6001 MQ493E1A2002 MQ493E1AC001 MQ493E1A1002 MQ493E19C002 MQ493E1A3003 MQ493E1A1004 MQ493E1A3002