FDA device recall Z-0867-2022 was initiated by Shimadzu Medical Systems on January 10, 2022 and is designated Class II. Reason for recall: It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation. The recall status is ongoing. Affected quantity: 25 systems.
Recall Details
- Product Type
- Devices
- Report Date
- April 13, 2022
- Initiation Date
- January 10, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 25 systems
Product Description
MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
Reason for Recall
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
Distribution Pattern
U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A
Code Information
Model: FLUOROspeed X1 Generator Number: D150-40 Serial Numbers: MQ493E19C001 MQ493E3B8002 MQ493E3B9001 MQ493E1B4001 MQ493E1A1003 MQ493E1A3001 MQ493E1B5001 MQ493E3B8001 MQ493E1A5001 MQ493E1A1005 MQ493E1A2001 MQ493E1A4001 MQ493E1B6002 MQ493E1A8001 MQ493E1B4002 MQ493E1A4003 MQ493E1A4002 MQ493E1B6001 MQ493E1A2002 MQ493E1AC001 MQ493E1A1002 MQ493E19C002 MQ493E1A3003 MQ493E1A1004 MQ493E1A3002