Z-0870-2020 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 5, 2020
- Initiation Date
- September 27, 2019
- Termination Date
- N/A
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 15
Product Description
EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths.
Reason for Recall
On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp.
Distribution Pattern
Worldwide Distribution.
Code Information
The defect concerns a total of fifteen (15) of our EQ-400 products. manufactured between June 3, 2016 and September 26, 2019.