Z-0870-2020 Class II Ongoing

Recalled by Energetiq Technology Inc — Woburn, MA

FDA device recall Z-0870-2020 was initiated by Energetiq Technology Inc on Sep 27, 2019 and is designated Class II. Reason for recall: On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted… The recall status is ongoing. Affected quantity: 15.

Recall details

Product Type
Devices
Report Date
Feb 5, 2020
Initiation Date
Sep 27, 2019
Voluntary/Mandated
FDA Mandated
Product Quantity
15

Product description

EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths.

Reason for recall

On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp.

Distribution pattern

Worldwide Distribution.

Code information

The defect concerns a total of fifteen (15) of our EQ-400 products. manufactured between June 3, 2016 and September 26, 2019.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0870-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Energetiq Technology Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Energetiq Technology Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Energetiq Technology Inc →

Source: openFDA record for Z-0870-2020. Report a data issue: hello@regdatalab.com.