Z-0870-2020 Class II Ongoing

Recalled by Energetiq Technology Inc — Woburn, MA

Recall Details

Product Type
Devices
Report Date
February 5, 2020
Initiation Date
September 27, 2019
Termination Date
N/A
Voluntary/Mandated
FDA Mandated
Product Quantity
15

Product Description

EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths.

Reason for Recall

On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp.

Distribution Pattern

Worldwide Distribution.

Code Information

The defect concerns a total of fifteen (15) of our EQ-400 products. manufactured between June 3, 2016 and September 26, 2019.