Z-0879-2013 Class II Terminated

Recalled by Baylis Medical Company Inc. — Mississauga, Ontario

FDA device recall Z-0879-2013 was initiated by Baylis Medical Company Inc. on Oct 4, 2012 and is designated Class II. Reason for recall: A report of an event was received indicating that upon opening a box containing 10 sterile cannula by a nurse, the pouch package of 5 of the 10 individually packaged cannula had not been sealed along… The recall status is terminated (terminated May 23, 2013). Affected quantity: 1780.

Recall details

Product Type
Devices
Report Date
Mar 6, 2013
Initiation Date
Oct 4, 2012
Termination Date
May 23, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1780

Product description

BMC RF Cannula Curved Sharp RadiOpaque, Non-Pyrogenic, Active Tip: 10mm, Gauge 16 Ga, Length: 100mm. The BMC RF Cannula is sealed in a Tyvek pouch (primary packaging) and then placed into a white shelf-box (secondary packaging) with the instructions for use. A predetermined number of white shelf-boxes are then placed into shipping cartons (tertiary packaging). The BMC RF Cannula is intended for use in radiofrequency heat lesion procedures for relief of pain.

Reason for recall

A report of an event was received indicating that upon opening a box containing 10 sterile cannula by a nurse, the pouch package of 5 of the 10 individually packaged cannula had not been sealed along one edge. Thus the pouch was open and the device could fall out of the pouch and on to the floor when retrieved from the shelf box. The cannulas were not used on any patients, as the unsealed packages

Distribution pattern

International distribution to the country of Australia.

Code information

Device Model #: 1 PMF 16-145-10CS. Lot number: PFFB 190811.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0879-2013
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Baylis Medical Company Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Baylis Medical Company Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Baylis Medical Company Inc. →

Source: openFDA record for Z-0879-2013. Report a data issue: hello@regdatalab.com.