Z-0879-2026 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0879-2026 was initiated by Siemens Medical Solutions USA, Inc on Oct 24, 2025 and is designated Class II. Reason for recall: There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary… The recall status is ongoing. Affected quantity: 1 unit (OUS only).

Recall details

Product Type
Devices
Report Date
Dec 10, 2025
Initiation Date
Oct 24, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit (OUS only)

Product description

MAMMOMAT Fusion;

Reason for recall

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AR, CA, NY and the countries of Australia, Canada, India.

Code information

Model Number: 10140000; UDI-DI: 04056869009063; System Serial Numbers: 10559;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0879-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0879-2026. Report a data issue: hello@regdatalab.com.