Z-0887-2022 Class II Ongoing

Recalled by Maquet Cardiovascular, LLC — Wayne, NJ

FDA device recall Z-0887-2022 was initiated by Maquet Cardiovascular, LLC on Feb 9, 2022 and is designated Class II. Reason for recall: Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses. The recall status is ongoing. Affected quantity: 60.

Recall Details

Product Type
Devices
Report Date
Apr 13, 2022
Initiation Date
Feb 9, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
60

Product Description

Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.

Reason for Recall

Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.

Distribution Pattern

Worldwide - United States (AL, AZ, CA, FL, IL, KY, LA, MO, MT, NM, OH, PA, TX, WI), Germany, Republic of Korea (South Korea), Russia, Thailand, United Arab Emirates.

Code Information

UDI# 00607567700901, Lots (3): 25153700, 25157489, 25158616. Please note affected lot numbers (25153700, 25157489, 25158616) appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number (200501), which is etched onto all 3 device components on side of handle and on underside above each pin.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0887-2022
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.