Z-0887-2022 Class II Ongoing
FDA device recall Z-0887-2022 was initiated by Maquet Cardiovascular, LLC on Feb 9, 2022 and is designated Class II. Reason for recall: Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses. The recall status is ongoing. Affected quantity: 60.
Recall Details
- Product Type
- Devices
- Report Date
- Apr 13, 2022
- Initiation Date
- Feb 9, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 60
Product Description
Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.
Reason for Recall
Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.
Distribution Pattern
Worldwide - United States (AL, AZ, CA, FL, IL, KY, LA, MO, MT, NM, OH, PA, TX, WI), Germany, Republic of Korea (South Korea), Russia, Thailand, United Arab Emirates.
Code Information
UDI# 00607567700901, Lots (3): 25153700, 25157489, 25158616. Please note affected lot numbers (25153700, 25157489, 25158616) appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number (200501), which is etched onto all 3 device components on side of handle and on underside above each pin.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0887-2022
- openFDA file date
- Oct 26, 2026
- Loaded by RegDataLab