Z-0894-2018 Class II Terminated

Recalled by Bausch & Lomb Inc Irb — Rochester, NY

Recall Details

Product Type
Devices
Report Date
March 14, 2018
Initiation Date
March 3, 2017
Termination Date
June 25, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,236,749 knives in total

Product Description

LASEREDGE Knives, 2.75MM THIN SLIT ANGLED BEVELED UP (6/BOX)(6/BOX), REF/PRODUCT CODE E7759A, STERILE, Rx Only Product Usage: The LaserEdge Knives are intended for ophthalmic surgery. There is no therapeutic use and they are available to surgeons for use in procedures.

Reason for Recall

The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing.

Distribution Pattern

US Nationwide Distribution

Code Information

none