Z-0918-2016 Class II Terminated

Recalled by Xeridiem Mediem Medical Devices Inc — Tucson, AZ

Recall Details

Product Type
Devices
Report Date
March 2, 2016
Initiation Date
December 23, 2015
Termination Date
October 25, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40

Product Description

EndoVive 3s Low Profile Balloon Kits Part Number: M00548820 (XMD P/N 70-0050-120) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach.

Reason for Recall

A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.

Distribution Pattern

Nationwide distribution to MA only.

Code Information

Part Number: M00548820 (XMD P/N 70-0050-120) Lot Numbers: 929724