Z-0926-2017 Class I Terminated
FDA device recall Z-0926-2017 was initiated by Nurse Assist, Inc on Oct 4, 2016 and is designated Class I. Reason for recall: Potential contamination with B. cepacia. The recall status is terminated (terminated Aug 16, 2017). Affected quantity: 386,175 syringes.
Recall details
- Product Type
- Devices
- Report Date
- Jan 11, 2017
- Initiation Date
- Oct 4, 2016
- Termination Date
- Aug 16, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 386,175 syringes
Product description
Normal Saline Flush 0.9% USP Sodium Chloride Injection Syringe, 3 ml fill, 5 ml fill and 10 ml fill. Product codes: 1203, 1205, 1210, 1210-BP Product usage: Saline flush
Reason for recall
Potential contamination with B. cepacia.
Distribution pattern
Worldwide Distribution - US (Nationwide), St. Thomas, US Virgin Islands, and Internationally to Canada.
Code information
Item Number 1203: 1511075, 1604273; Item Number 1205: 1511001, 1602028, 1603060; Item Number 1210: 1509282, 1511328, 1511329, 1512048, 1512053, 1512072, 1601355, 1602048, 1602137, 1602178, 1603029, 1605387, 1606076; Item Number 1210-BP: 1604431,1604432, 1604433, 1606254, 1607018, 1607195, 1607418 (All lots within expiry)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0926-2017
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
Need the full FDA record for Nurse Assist, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.