Z-0929-2018 Class II Terminated
Recalled by Euro Diagnostica AB
— Malmo, N/A
FDA device recall Z-0929-2018 was initiated by Euro Diagnostica AB on Oct 15, 2015 and is designated Class II. Reason for recall: Complaint investigation concluded that although product quality requirements stated in the instructions for use were met at lot release, the high background in the ELISA plate, the strong IgM… The recall status is terminated (terminated Mar 29, 2019). Affected quantity: 1090.
Recall details
- Product Type
- Devices
- Report Date
- Mar 14, 2018
- Initiation Date
- Oct 15, 2015
- Termination Date
- Mar 29, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1090
Product description
DIASTAT anti-Nuclear antibody (ANA) / DIASTAT ANA ELISA. Catalog Number: FANA200. Product Usage: The DIASTAT¿ anti-nuclear antibody (ANA) test is a qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of ANAs in human serum or EDTA plasma. It detects ANAs against Sm, Sm/RNP, Ro (SS-A), La (SS-B), Scl- 70, Jo-1, dsDNA, histone and centromere antigens.The test may be used to screen out samples which are negative for all ANAs. Samples that give a positive test result should be further tested to identify the antigen-specific antibody or antibodies present. Quantitative/qualitative ELISA kits are available for the individual detection of Sm, Sm/RNP, Ro (SS-A), La (SS-B), dsDNA and centromere ANAs. Qualitative kits are available for the individual detection of Scl-70 and Jo-1. ANA detection represents one parameter in a multicriterion diagnostic process.
Reason for recall
Complaint investigation concluded that although product quality requirements stated in the instructions for use were met at lot release, the high background in the ELISA plate, the strong IgM conjugate, and the low reference control led to an increased risk of equivocal and/or false positive sample test results for the two FANA200 kit lots SS 1500 and SS 2009.
Distribution pattern
Worldwide Distribution and US. Nationwide
Code information
UDI: 07340058410486. Lot numbers: SS 1500 and SS 2009.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0929-2018
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Euro Diagnostica AB? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Euro Diagnostica AB: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Euro Diagnostica AB →
Use it in ChatGPT
Get a free API key
How the connector works