Z-0932-2019 Class II Terminated

Recalled by Philips Medical Systems Nederlands — Best

Recall Details

Product Type
Devices
Report Date
March 6, 2019
Initiation Date
December 21, 2018
Termination Date
November 19, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11226 total

Product Description

Multiva 1.5T, Model Number 781076. Nuclear Magnetic Resonance Imaging System

Reason for Recall

There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.

Distribution Pattern

U.S. Nationwide distribution.

Code Information

All systems