Z-0940-2026 Class II Ongoing

Recalled by PATH — Germering

FDA device recall Z-0940-2026 was initiated by PATH on November 6, 2025 and is designated Class II. Reason for recall: Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These di… The recall status is ongoing.

Recall Details

Product Type
Devices
Report Date
December 24, 2025
Initiation Date
November 6, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity

Product Description

MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)

Reason for Recall

Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.

Distribution Pattern

Domestic: WI. International: Ireland.

Code Information

Part Number: 8-69-41100; UDI-DI: 04260223141355; Serial Numbers: 1008852, 1008888, 1008916, 1008925, 1008935, 1008945, 1008951, 1008957, 1008960, 1008962, 1008966, 1008967, 1008987, 1008990, 1009054, 1009065, 1009066, 1009131, 1009137, 1009138, 1009139, 1009145, 1009148, 1009155, 1009159, 1009161, 1009169, 1009184, 1009188, 1009191, 1009305, 1009315, 1009316, 1009319, 1009321, 1009322, 1009325, 1009328, 1009343, 1009352, 1009356, 1009357, 1009359, 1009366, 1009372, 1009375, 1009376, 1009386, 1009388, 1009390, 1009396, 1009399, 1009400, 1009404, 1009411, 1009415, 1009419, 1009426, 1009427, 1009432, 1009439, 1009442, 1009454, 1009456, 1009460, 1009463, 1009464, 1009468, 1009475, 1009481, 1009488, 1009494, 1009495, 1009497, 1009848, 1009855.