Z-0945-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- February 10, 2021
- Initiation Date
- December 28, 2020
- Termination Date
- September 14, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 52,000
Product Description
Tourni-Cot - Universal, UDI 00855364004085, Model Number TCU, Non-pneumatic Tourniquet - Product Usage: The device is intended to be used in surgery on fingers or toes to provide a blood-less field.
Reason for Recall
A failure can occur if the user attempts to remove the device by partially snipping the edge and pulling on it. The device may tear into two pieces with a portion remaining in place, which can continue to constrict the vessels. This type of failure can result in digit ischemia.
Distribution Pattern
Worldwide distribution - US Nationwide distribution and the countries of Australia, Canada, France, Germany, Greece, Italy, Japan, Netherlands, Singapore, Spain, Sweden, Switzerland, and United Kingdom.
Code Information
All lots