Z-0949-2020 Class I Terminated

Recalled by Philips Respironics, Inc. — Murrysville, PA

Recall Details

Product Type
Devices
Report Date
February 26, 2020
Initiation Date
November 19, 2019
Termination Date
March 31, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

Trilogy EVO Ventilator, France - Model Number: FR2110X14B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories.

Reason for Recall

Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.

Distribution Pattern

US: None OUS: CH DE DK IT FR IE GB KW QA PL ES AT PH AE SI CZ

Code Information

Model/Material Number: FR2110X14B Serial Numbers: H248213027F13, H249718745CF4, H252330869C7A, H252331053AFD, H255249974881,