Z-0998-2019 Class II Terminated

Recalled by Ossur Americas — Foothill Ranch, CA

Recall Details

Product Type
Devices
Report Date
March 20, 2019
Initiation Date
June 21, 2018
Termination Date
September 4, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
824

Product Description

Ossur Skull Pin, Sterile Ceramic Tip Pin, Part/ Description: 516CS/ Skull Pin 2-1/4in CT Sterl 1PK; 516DS/ Skull Pin 3in CT Sterl 1PK; 517CS/ Skull Pin 2-1/4in CT Sterl 4PK; 517DS/ Skull Pin 3in CT Sterl 4 PK; 522CS/ Skull Pin 2-1/4in CT Sterl 6PK; 522C/ SKULL PIN 2-1/4 CT 6 PK

Reason for Recall

Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.

Distribution Pattern

U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium

Code Information

All lots of the products manufactured from 06/30/2009 - 12/13/2018