Z-1001-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- February 12, 2020
- Initiation Date
- January 2, 2020
- Termination Date
- November 3, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 261
Product Description
Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011
Reason for Recall
A mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle.
Distribution Pattern
Worldwide distribution. US Nationwide, Venezuela, Africa, Italy, Brazil, and Ukraine. No governmental, Canadian, or Mexican consignees.
Code Information
Part Number 1400075 (US) Part Number 1400011 (OUS) All serial numbers distributed prior to 3/21/2019