Z-1016-2015 Class II Terminated

Recalled by Mobius Imaging, LLC — Ayer, MA

FDA device recall Z-1016-2015 was initiated by Mobius Imaging, LLC on Dec 8, 2014 and is designated Class II. Reason for recall: There is a risk that during the transfer of an image and navigation data to the Brainlab Curve Image Guided Surgery Navigation System after a CT scan, an error may occur, causing either no navigation… The recall status is terminated (terminated Jul 6, 2015). Affected quantity: 21 devices distributed. 15 of these devices affected by the software v. 1.1.1+ patch.

Recall details

Product Type
Devices
Report Date
Feb 4, 2015
Initiation Date
Dec 8, 2014
Termination Date
Jul 6, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21 devices distributed. 15 of these devices affected by the software v. 1.1.1+ patch

Product description

Airo Mobile Intraoperative CT I Airo; Model #: MobiCT-32. Computed Tomography x-ray system.

Reason for recall

There is a risk that during the transfer of an image and navigation data to the Brainlab Curve Image Guided Surgery Navigation System after a CT scan, an error may occur, causing either no navigation data or incorrect navigation data being transferred to the curve.

Distribution pattern

USA including AK, IL, WA, MA, NY, WV, NC, TX, CA, PA and Internationally to Germany, Netherlands, Belgium, Switzerland, Saudi Arabia, Egypt.

Code information

Serial #s: 2251 (#1)*, 2324 (#2), 2328(#3), 2344 (#4), 2346 (#5), AIRO-0106, AIRO-0107*, AIRO-0108, AIRO-0110, AIRO-0111, AIRO-0112, AIRO-0113, AIRO-0114, AIRO-0115, AIRO-0116, AIRO-0117, AIRO-0118, AIRO-0119, AIRO-0120, AIRO-0122, and AIRO-0123. *Demo system

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1016-2015
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1016-2015. Report a data issue: hello@regdatalab.com.