Z-1022-2013 Class II Terminated

Recalled by Genetic Testing Institute,Inc — Waukesha, WI

FDA device recall Z-1022-2013 was initiated by Genetic Testing Institute,Inc on Feb 4, 2013 and is designated Class II. Reason for recall: Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in… The recall status is terminated (terminated Mar 21, 2016). Affected quantity: 386 kits (370 US, 16 OUS).

Recall details

Product Type
Devices
Report Date
Apr 3, 2013
Initiation Date
Feb 4, 2013
Termination Date
Mar 21, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
386 kits (370 US, 16 OUS)

Product description

Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA).

Reason for recall

Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an invalid assay is obtained, those results should not be reported and the test should be re-run. An invalid assay would not result in significant harm to a pati

Distribution pattern

Worldwide Distribution - USA (nationwide) and Internationally to Canada, China, Slovakia, and Spain.

Code information

Kit Lot Number 3000389

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1022-2013
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Genetic Testing Institute,Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1022-2013. Report a data issue: hello@regdatalab.com.