Z-1022-2013 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 3, 2013
- Initiation Date
- February 4, 2013
- Termination Date
- March 21, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 386 kits (370 US, 16 OUS)
Product Description
Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA).
Reason for Recall
Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an invalid assay is obtained, those results should not be reported and the test should be re-run. An invalid assay would not result in significant harm to a pati
Distribution Pattern
Worldwide Distribution - USA (nationwide) and Internationally to Canada, China, Slovakia, and Spain.
Code Information
Kit Lot Number 3000389